Hemodialysis Machine ADM-8
Dialysate
Dialysate temperature: Preset range 33.0℃~39.0℃, ±0.1℃
Dialysate flow: Less than 800ml/min (Adjustable)
Concentration of dialysate: 12.1 mS/cm~16.0 mS/cm, ±0.1 mS/cm
Dialysate ratio: Various proportions
UF (dehydration): 0 ml/h~4000 ml/h
Resolution: 1ml
Precision: ±1%
Heparin Pump
Syringe size: 20, 30, 50ml
Flow range: 0 ml/h~10 ml/h
First dose range: 1.0~20.0ml
Resolution: 0.1ml
Precision: ±0.2 ml or ±5% of the reading
Extracorporeal Circulation
Venous pressure: -180 mmHg ~ +600 mmHg, ±10 mmHg
Arterial pressure: -380 mmHg ~ +400 mmHg, ±10 mmHg
TMP pressure: -180 mmHg ~ +600 mmHg, ±20 mmHg
Blood pump, assistant pump flow range: 30 ml/min ~ 600 ml/min (Inner diameter Ф8mm)
Resolution: 1ml
Precision: ±10ml/min or 10% of the reading
Blood leak detection: Photo chromic detecting
Bubble detection: Double protection
Disinfection and Cleaning Procedures
Chemical decalcification: Room temperature/85℃, flow: 300 ~ 800ml/min (adjustable)
Chemical disinfection: Room temperature, flow: 300 ~ 800ml/min (adjustable)
Heat disinfection: 85℃, flow: 300 ~ 800ml/min (adjustable)
Intelligent disinfection setting from Monday to Sunday makes disinfection safer and more reliable.
Technology Data
Weight: 66kg
Input power: 1500w
Standby battery: 30 minutes
Water inlet pressure: 0.15 MPa ~ 0.6 MPa
Water inlet temperature: 5℃ ~ 30℃
Working environment: Temperature 10℃ ~ 35℃ at relative humidity of less than 70%
There are no reviews yet.